List-Magnetik GmbH

Magnet permeability meter Ferromaster

magnet_permeability_meter_ferromaster.jpg

With the Magnet permeability meter List-Magnetik Ferromaster you can easily determine the relative magnetic permeability µr of feebly magnetic materials and workpieces with a permeability between 1.001 and 1.999.

The permeability is measured by touching the workpiece with the probe tip and reading the result from the display.

The magnetic permeability or magnetic conductivity makes a statement as to how strong a material is magnetizable.

Typical applications are: non-destructive testing of materials, e. g. quality control of stainless steel, material selection for electron-/ion-beam equipment, detection of material defects induced by mechanical or thermal stress. Standard-compliant permeability measurements according to ASTM A342 Test Method 4 and EN 60404-15 Procedure 6 can be carried out with the Ferromaster.

The probe head contains a permanent magnet that magnetizes the sample near the probe tip. Two sensitive magnetic field sensors in differential circuit measure the field distortions caused by the magnetization of the sample. The device is calibrated to precisely calibrated reference standards from the National Physical Laboratory (NPL, Teddington, UK). Recalibration is easy. A calibration standard is included with every device.

As a special feature, the Ferromaster is equipped with a robust, waterproof hand-held housing (protection class IP65) and is therefore ideal for use in harsh industrial environments. The built-in battery enables approx. 50 hours of operation.

The magnetic permeability of a material generally depends heavily on the strength and frequency of the magnetizing field. The probe head of the Ferromaster contains a permanent magnet that generates a constant magnetic field of around 35 kA/m at the test probe.

The measured value of the permeability depends on the dimensions of small samples. The sensitivity of the device takes z. B. with the thickness of the sample.
From a sample thickness of around 5 mm and a lateral extent of around 2 cm, the measurement is independent of the size.

Technical Data Magnet permeability meter Ferromaster

Measurement range:µr = 1.000 to 1.999
Resolution:0.001
Display:LCD 3½ digits
Zeroing:automatically
Accuracy at 20 °C:(µr - 1) x 5%
Operating temperature:0 to 50 °C
Field strength at probe tip:~ 35 kA/m
Minimum measurement sample size:20x20x8 mm (LxWxH)
Power supply:9 V PP3 battery
Operation time:~ 50 h
Environmental protection:IP65
Standards:Permeability measurement compliant with ASTM A342 and EN 60404-15
Traceability:Calibrated to reference standards of National Physical Laboratory, UK
Dimensions of unit:151 mm x 82 mm x 33 mm
Length of probe:35 mm
Diameter of probe:19 mm
Length of connection cable:1.5 m
Weight of complete instrument:280 g